Verified Peptide Test Results

Anonymized results from peptides submitted to our lab, showing what the supplier claimed next to what the instrument measured. Every row came off a signed Certificate of Analysis.

Last reviewed

Claimed vs measured: sample results

Supplier-claimed purity compared with measured HPLC purity and LC-MS identity for independently tested peptide samples
PeptideClaimedMeasuredMethodResultFinding
Semaglutide99.0%94.2%RP-HPLC 214 nmBelow claimTwo deletion-sequence impurities above 1.5% area.
Tirzepatide99.5%98.9%RP-HPLC + LC-MSMatchObserved mass within 0.4 Da of theoretical.
Retatrutide99.0%88.6%RP-HPLC + LC-MSBelow claimLong sequence; truncated fragment at 6.1% area.
BPC-15799.0%97.8%RP-HPLC 214 nmMatchWithin typical research-grade tolerance.
TB-500 (TB-4 fragment)98.0%71.4%RP-HPLC + LC-MSBelow claimMannitol excipient counted as active in supplier COA.
Tesamorelin99.0%96.1%RP-HPLC + LC-MSBelow claimOxidized methionine variant at 2.2%.
"GLP-1 blend" vial99.0%n/aLC-MS/MS identityIdentity mismatchObserved mass matched liraglutide, not the labeled sequence.
MOTS-c98.0%98.3%RP-HPLC + ESI-MSMatchIdentity and purity both confirmed.

Supplier names and batch numbers are withheld under client confidentiality. Methods, claimed values, and measured values are reported exactly as issued.

What the aggregate data says

73%

of samples showed a purity discrepancy against the supplier claim

8.4 pts

average gap between claimed and measured HPLC purity

27%

of supplier COAs matched independent results line for line

The full statistical breakdown, including contaminant and endotoxin failure rates, lives on our peptide testing statistics page. For how these numbers are produced, see HPLC peptide purity testing and mass spectrometry identity confirmation.

Get your own batch verified

Ship a sealed vial and we return a batch-specific Certificate of Analysis with the raw chromatogram and mass spectrum attached. Independent lab, no supplier affiliation, 9-11 business day turnaround.

Related reading

Methodology: purity by reverse-phase HPLC with UV detection at 214 nm, area percent of all integrated peaks. Identity by LC-MS/MS or ESI-MS against the theoretical monoisotopic mass. Heavy metals by ICP-MS per USP <232>/<233>. Endotoxin by kinetic chromogenic LAL per USP <85>.

Verified results: frequently asked questions